We use Peltier Device for Thermal Test Solutions. It’s simple, compact, easy to use, accurate & less maintenance is required.
Variety of applications, e.g. Semiconductor device, Automotive, Military, Aerospace, electronic parts, and the materials tests, etc.
The system requires:
Max Power Peltier Thermal Head, iT550 or iT555
System Controller FC-4610
Chiller CCA-1111, Nitrogen Generator TWC-05
Simulate temperature environment test:
From-55°C to+135°C, temperature control accuracy ± 0.1°C.
4 min. from ambient to -55°C
3 min. from -55°C to+135°C
Equippedwith RS remote control, temperature programmable controller.
MAX POWER THERMAL HEAD
iT550/iT555
iT550: the cooling water joints are under (right photo above)
iT555: the cooling water joints are side (left photo above).
Specifications
Model
iT550
iT555
Temperature Range (˚C)
-55˚C to +135˚C
Temperature control accuracy (˚C)
±0.1˚C
Temperature sensor
Pt100 Platinum Measurement Temperature Resistance
Heat radiation method
Water-cooled (use cooling chiller)
Cooling heating method
Use Cascade Peltier device type UT-4300
Driving voltage (V) /current (A)
DC15V / 6A
Contact Surface Dimension (mm)
30×30mm or 20×20mm (Standard)
Weight (g)
196g
SYSTEM CONTROLLER
FC-4610
FC-4610 is a temperature controller developed to drive the Cascade Peltier Device Thermal Head (iT550 or iT555). The thermo regulator is operated by a programmable controller. Highly accurate temperature control (± 0.1˚C), wide temperature range -60˚C to +150˚C (temperature setting range) and can interface with PC (option). Peltier temperature adjustment controller (FC 6410) is equipped with pt100 sensor for temperature measurement.
Specifications
Model
FC-4610
Temperature range (˚C )
-60˚C to +150˚C
Temperature control accuracy (˚C)
±0.1˚C
Temperature sensor
Pt100 Platinum Measurement temperature resistance
Use For Peltier
Output voltage (V) /current (A)
Use for Cascade Peltier device only. DC 12V・4V/MAX 6A
Control method /temperature regulator
Proportion time PID control/ SDC-35 (Yamatake Corp.)
Temperature program
8 point max.
Power-supply voltage (V)
AC100 to 230
Dimension (mm)
320×220×120
Serial port/PC port
No/ Yes
Temperature Characteristic
CHILLER/NITROGEN GENERATOR
Cooling Chiller CCA-1111
The cooling surface of -55˚C is achieved by circulating cold anti-freeze in the Peltier device.
Nitrogen Generator TWC-05
Dew is prevented on the surfave of the Peltier device by nitrogen gas when the temperature is below zero.